Model No.
Parameters |
Enteral Feeding Pump |
Flow rate range(ml/h) |
1~400, step by 1ml/h |
Accuracy of flow rate |
±10%, with dedicated IV set |
volume limit(ml) |
0-9999, step by 1 ml |
Accuracy of infusion volume |
±10%, with dedicated IV set |
Prime rate and accuracy(ml/h) |
1000 ±10% |
Flush rate and accuracy(ml/h) |
1000 ±10% |
One-time flush volume(ml) |
40±10% |
Max infusion pressure |
≥0.1 MPa |
Occlusion alarm value(pressure) |
50KPa±10KPa |
IV set |
Refer to the recommended IV set list |
Bubble sensitivity |
Detestability of air bubble≥50uL |
Alarming |
Occlusion, Battery exhausted |
Indicating functions |
Bubble in line, VTBI completion, Flush completion, Low battery, Malfunction(ER series),Pause overtime, Power interruption, Infusion mode, Working state |
Noise |
With equipment operating on a wooden table in the maximum flow rate, and in the distance of 1 meter in every direction, the noise level is below 50dB. |
Power |
AC:100V-240V 50 Hz /60Hz |
Internal power |
After fully charged, internal DC11.1V rechargeable Li battery can run over 5 hours at intermediate rate |
Weight |
2.5kg |
Classification |
Type BF, Class II continuous volumetric feeding pump,IPX1(sealed drip-proof device) |
Size |
170mm(W)×155mm(L)×180mm(H) |
Optional Accessory |
Fluid warmer, Dedicated disposable nutrition IV set |